RT2-0.5-80 Stationary X-Ray Tube is especially designated for bone densimeter x-ray system and available for nominal tube voltage constant potential circuit
RT2-0.5-80 tube has one focus.
The integrated high quality tube with glass design has one super imposed focal spot and areinforced anode. The high anode heat storage capacity ensures a wide range of applications for bone densimeter x-ray application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.
RT2-0.5-80 Stationary X-Ray Tube is especially designated for bone densimeter x-ray system and available for nominal tube voltage constant potential circuit.
This tube is especially designed for Bone Mineral Densitometer x-ray system unit. Nominal Tube Voltage(self-rectified)……..………………..….………..…….……………95kV (constant potential)………………..….………..…….……………...................................80kV Nominal Tube Current .……………………………..….……….…………………………........2mA Nominal Operating Condition………………………….…….....................80kV@2mA@2.4V Nominal Anode Cooling Rate……………………………………………………………….....160W Anode Heat storage Capacity ………………………………………………………………......16kJ Filament current………………………………………………………….….....1.95 A@2.4V±0.2V Focal spot …………………………………………………………………......0.5 (IEC60336/2005) Target Angle………………………………………………………………….…….…………..............12° Effective X-ray Cone Beam……………………………………………………….…….......24°X90° Target Material………………………………………………………………….……….........Tungsten Cathode type……………………….…………………………………………........Tungsten filament Dimensions ……………………………………………..….....125mm length by 35mm diameter Weight…………………………………………………………………………............approx.145 grams
Maintaining seasoning schedule
Prior to usage, season the tube in accordance with the seasoning schedule given below until
the required tube voltage is attained. Example given – needs to revise by the manufacturer
and specified in the data sheet of the part:
Initial incoming seasoning and seasoning schedule for idle period (more than 6 months)
Circuit: DC (Center grounded)
When the tube current is unstable in seasoning, immediately switch off the tube voltage and
after an interval of 5 minutes or more, increase the tube voltage gradually from the low
voltage while making sure that the tube current is stable.
The withstand voltage performance of the tube unit will be lowered as the exposure time and
the number of operation increase. Stain-like impact traces might appear on the x-ray tube
target surface by slight discharge in the course of the seasoning. These phenomena are one
process to recover the withstanding voltage performance at that time.
Therefore, if it is in stable operation at the maximum tube voltage of seasoning subsequent
to them, the tube unit can be used without any interference to its electrical performance
which in use.
Elevated anode heat storage capacity and cooling
Constant high dose yield
Excellent lifetime