This high voltage cable’s typical applications are as followings:
1、Mammography and other scientific X-ray, electron beam or laser
equipment
2、Low power high voltage testing and measuring equipment.
1. High flexibility
2. Small diameter
3. 95% braided shielding density
4. The rated voltage of the cable is 60kVDC
Number of conductor | 1 |
Rated voltage | 60kVDC |
Routine test voltage (high voltage insulation) | 90kVDC/10min |
Rated conductor current | 31A |
Nominal outside diameter | 12.4mm±0.5mm |
Thickness of PVC jacket | 1.0mm |
Thickness of high voltage insulation | 2.9mm |
Diameter of core-assembly | 1.8mm |
Insulation resistance core to shield @20℃ | ≥1×1012Ω·m |
DC resistance of conductor at 20℃ | 8.9±0.45Ω /km |
Shield resistance@20℃ | 8.0±0.45Ω /km |
Max Capacitance between conductor and shield | 120±12pF/m |
Cable Min bending radius(static insulation) | 22mm |
Cable Min bending radius(dynamic installation) | 45mm |
Operating temperature | -10℃~+70℃ |
Storage temperature | -40℃~+70℃ |
Net weight | 206.8kg/km |
Number of conductor | 1 |
Rated voltage | 60kVDC |
Routine test voltage( between pin and ground) | 75kVDC/15min |
Maximum rated current | 25A |
Maximum continuous working temperature of plug shell | 100℃ |