This high voltage cable’s typical applications are as followings:
1、Mammography and other scientific X-ray, electron beam or laser
equipment
2、Low power high voltage testing and measuring equipment.
1. High flexibility
2. Small diameter
3. 95% braided shielding density
4. The rated voltage of the cable is 60kVDC
| Number of conductor | 1 | 
| Rated voltage | 60kVDC | 
| Routine test voltage (high voltage insulation) | 90kVDC/10min | 
| Rated conductor current | 31A | 
| Nominal outside diameter | 12.4mm±0.5mm | 
| Thickness of PVC jacket | 1.0mm | 
| Thickness of high voltage insulation | 2.9mm | 
| Diameter of core-assembly | 1.8mm | 
| Insulation resistance core to shield @20℃ | ≥1×1012Ω·m | 
| DC resistance of conductor at 20℃ | 8.9±0.45Ω /km | 
| Shield resistance@20℃ | 8.0±0.45Ω /km | 
| Max Capacitance between conductor and shield | 120±12pF/m | 
| Cable Min bending radius(static insulation) | 22mm | 
| Cable Min bending radius(dynamic installation) | 45mm | 
| Operating temperature | -10℃~+70℃ | 
| Storage temperature | -40℃~+70℃ | 
| Net weight | 206.8kg/km | 
| Number of conductor | 1 | 
| Rated voltage | 60kVDC | 
| Routine test voltage( between pin and ground) | 75kVDC/15min | 
| Maximum rated current | 25A | 
| Maximum continuous working temperature of plug shell | 100℃ |