Home Categories Send inquiry

Mobile X-ray Tube Cei OX110-5

KL25-0.6/1.5-110 stationary anode x-ray tube is especially designed for C-arm equipment and available for nominal tube voltage with high frequency or DC generator.
The integrated high quality tube with glass design has two super imposed focal spot and a reinforced anode. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.

KL25-0.6/1.5-110 is especially designed for C-arm equipment and available for nominal tube voltage with DC generator.

This tube has focus 1.5 and 0.6 foci and is available for maximum tube voltage 110 kV.

 

Nominal Tube Voltage 110kV
Nominal Focal Spot small:0.6   large:1.5(IEC60336/2005)
Filament Characteristics small:Ifmax=4.5A,Uf=5±0.5  large:Ifmax=4.5A,Uf=6.3±0.8V
Nominal Input Power(at 1.0s) small:spot0.6kW  large:spot3.5kW
Maximum Continuous Rating 225W
Anode Heat Storage Capacity 30kJ
Target Angle 12°
Target Material Tungsten
Inherent Filtration Min 0.6mmAl equivalent at 75kV
Weight approx540g

Environmental Limits

Operating Limits (in dielectric oil):

Oil Temperature ................................................................................................10 ~ 60 °C

Oil Pressure ................................................................................................. 70 ~ 106 kPa

Shipping and Storage Limits: Temperature.................................................-40~ 70 °C

Humidity ........................................................................................................... 10 ~ 90 %

(No condensation)

Atmospheric Pressure ................................................................................. 50 ~ 106 kPa

Elevated anode heat storage capacity and cooling
Constant high dose yield
Excellent lifetime